Purpose The aim of this study is to present and validate the semi-automated MuscleExpert software against the widely used ImageJ by comparing muscle thickness (MT), muscle area (MA), and muscle quality (MQ-measured as grayscale).
Methods Ten volunteers participated in this study. Ultrasound images of the tibialis anterior and medial gastrocnemius muscles were acquired using 2D B-mode ultrasonography and analyzed on both software. Equivalence of means was evaluated via the two‐ one‐sided tests (TOST) procedure, the reliability between means was quantified by an intraclass correlation coefficient (ICC), the precision was expressed as the coefficient of variation as well as standard error of measurement (SEM), and the agreement across the measurement was assessed using Bland–Altman plots.
Results The TOST for MT, MA, and MQ was within their respective equivalence margins. The ICC was excellent for all measures (MT: ICC=0.999; MA: ICC=0.999; MQ: ICC=1.000). Precision was similarly high, with CVs of 0.49% for MT, 0.54% for MA, and 0.33% for MQ, demonstrating minimal variability between the two methods. SEM values were low across all outcomes, corresponding to 0.49 % for MT, 0.53% for MA, and 0.23% for MQ. Finally, the Bland–Altman analysis demonstrated minimal systematic differences and no proportional bias between ImageJ and MuscleExpert for all three metrics.
Conclusion The present findings indicate that MuscleExpert and ImageJ produce nearly identical results for MT, MA, and MQ. These findings support the integration of MuscleExpert into clinical and research workflows, offering a more efficient solution for muscle assessment without compromising measurement integrity.
Competing Interest StatementThe authors have declared no competing interest.
Funding StatementNo funding
Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The work was approved by the University of Maryland, Baltimore, ethics committee (number: HP-00109908)
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data AvailabilityData is available upon reasonable request.
Comments (0)