The number of people diagnosed with allergic diseases has been steadily increasing as modern environmental exposures evolve.1 In clinical practice, the diagnosis of allergies relies heavily on patient-reported symptoms, medical history, and physical examination findings to identify potential causative agents. To assess Immunoglobulin E sensitization, allergen-specific Immunoglobulin E (sIgE) levels can be measured using both in vivo and in vitro methods. Skin prick testing has traditionally been used, but may be limited in patients who cannot discontinue antihistamines or who have dermatologic conditions, and positive results may also occur in individuals without clinically relevant allergy.2 In vitro sIgE assays provide complementary information to skin testing and are widely used as screening tools for sensitization in clinical practice.3
Among the commercially available in vitro methods, multiple allergen simultaneous tests (MASTs) require only a small volume of serum, offer short processing times, and continue to evolve by expanding the allergen spectrum included in each panel.4 The prevalence and spectrum of allergic sensitization vary according to region, age, and ethnicity, underscoring the importance of selecting an appropriate MAST panel based on the characteristics of the target population. However, differences in allergen sources, extraction methods, solid-phase attachment techniques, and detection technologies mean that no single analytical method has yet been universally accepted as a reference standard.5
In this context, a comparative evaluation of different MAST systems is clinically meaningful, as it provides evidence regarding their analytical agreement, diagnostic reliability, and coverage of clinically relevant allergens. Such assessments can guide laboratories in selecting the most appropriate system for their local patient populations and national healthcare needs. In this retrospective study, we compared the diagnostic performances of four fully automated MAST analyzers commonly used in Korea.
Materials and Methods Study ParticipantsFrom November 2023 to December 2023, 200 residual serum samples were randomly collected from patients who underwent inhalant or food MAST panel testing for suspected allergic diseases at a single tertiary hospital. This study was approved by the Institutional Review Board of Severance Hospital (IRB No. 2023–3201-002).
In vitro sIgE MeasurementsAll serum samples were first tested using AdvanSure Alloscreen (Invitros, Cheongju-si, Korea), which served as the institutional reference method. Residual serum aliquots were subsequently analyzed on the same day using three additional systems: AdvanSure Alloscreen Max108 (Invitros, Cheongju-si, Korea), SGTi-allergy screen (Sugentech, Cheongju-si, Korea), and PROTIA allergy-Q (Protia, Seoul, Korea). All assays were performed according to the manufacturer’s instructions. The results were expressed on a semi-quantitative scale ranging from class 0 to 6 for all analyzers. The detailed technical specifications of each MAST system are listed in Table 1. For discrepant cases with sufficient residual serum, additional testing was performed using ImmunoCAP (Phadia AB, Uppsala, Sweden), which is considered the gold standard for sIgE measurements. A complete list of the allergens included in each panel is provided in Table S1.
Table 1 Technical Specifications and Distinguishing Features of the Four MAST Systems Evaluated
Statistical AnalysisAllergens were pre-specified into eight groups: dust, animal epithelium, weed, mold, tree, grass, insects, and food. The results from the AdvanSure Alloscreen served as the institutional reference for comparison, whereas ImmunoCAP results, when available, were considered the true reference standard.6 For allergens not included in the AdvanSure Alloscreen assay, concordant results across other assays were used as surrogate reference standards. Following Rim et al,4 positivity was defined as class ≥2 Pairwise analyses were restricted to 91 allergens shared across systems, while allergens unique to a single system were excluded (Table S1). For the class discrepancy analyses, allergens with <5 paired observations were excluded a priori.
The total agreement percentage was calculated as follows.7
The agreement levels were categorized as almost perfect (90–100%), substantial (80–90%), moderate (70–80%), fair (60–70%), or poor (< 60%). Cohen’s kappa (κ) coefficient was used to evaluate agreement between paired assays,8 with values interpreted as almost perfect (0.8–1.0), substantial (0.6–0.8), moderate (0.4–0.6), fair (0.2–0.4), and poor (<0.2). All statistical analyses were performed using Microsoft Excel 2016 (Microsoft, Redmond, WA, USA).
Results Study PopulationThe baseline characteristics of the study population are summarized in Table 2. A total of 200 residual serum samples were analyzed. Due to limited residual serum volume, the SGTi-Allergy Screen was performed on 186 samples, while the remaining three MAST systems were tested on all 200 samples. The median patient age was 21.5 years (range, 2–85 years). Among the 200 patients, 86 were pediatric patients (<18 years), including 54 males. Overall, males accounted for 121 of 200 (60.5%) patients.
Table 2 Baseline Characteristics of the Study Population and Sample Overview
Inter-System AgreementAcross all paired comparisons, the overall agreement among the MAST systems ranged from 91% to 93%, with Cohen’s κ values between 0.67 and 0.76, indicating substantial to almost perfect agreement (Table 3). The highest agreement was observed between AdvanSure Alloscreen and AdvanSure Alloscreen Max108 (93% agreement, κ = 0.76). For prevalent inhalant allergens in Korea, Dermatophagoides pteronyssinus (d1) and D. farinae (d2) showed inter-system agreement of 72–90% and 82–93%, respectively, with corresponding κ values of 0.47–0.79 and 0.60–0.82 (Table 3).9,10
Table 3 Pairwise Agreement and Cohen’s Kappa Values for Representative Inhalant and Food Allergens Across Four MAST Systems
Semi-Quantitative Class DifferencesInter-assay evaluations limited to overlapping allergens showed absolute ordinal class discrepancies generally ≤2. The most pronounced divergences occurred in the following food allergen categories: citrus mix (f33) (AdvanSure Alloscreen vs. AdvanSure Alloscreen Max108, comparison 1) and mussel (f37) (SGTi-Allergy Screen vs. PROTIA Allergy-Q, comparison 4), each demonstrating a three-class deviation. Figure 1 presents the distribution of these class discrepancies. In addition, mango (f91) showed an average 2.5-class discrepancy in AdvanSure Alloscreen vs. SGTi-Allergy Screen (comparison 2) and SGTi-Allergy Screen vs. PROTIA Allergy-Q (comparison 4).
Figure 1 Heatmap of class difference between MAST systems. Heat map showing absolute semi-quantitative class differences between paired multiple allergen simultaneous test (MAST) systems for overlapping allergens. Rows represent individual allergens, grouped by allergen category (dust, animal epithelium, weed, mold, tree, grass, and food), and columns represent pairwise system comparisons. Allergen codes correspond to the WHO/IUIS allergen nomenclature and are detailed in Table S1. Shading was applied only when class differences exceeded 1, with the color intensity increasing in 0.5-class increments; darker shading indicated larger discrepancies. White cells indicate class differences ≤1 or identical class results. Column headers (1–6) correspond to: 1. AdvanSure Alloscreen vs. AdvanSure Alloscreen Max108; 2. AdvanSure Alloscreen vs. SGTi-allergy screen; 3. AdvanSure Alloscreen vs. Protia allergy-Q; 4. SGTi-allergy screen vs Protia allergy-Q; 5. SGTi-allergy screen vs AdvanSure Alloscreen Max108; and 6. Protia allergy-Q vs AdvanSure Alloscreen Max108.
ImmunoCAP Verification of Discrepant ResultsDiscrepant samples with sufficient residual serum (n = 78; 31 allergens) underwent confirmatory testing using ImmunoCAP. Sensitization to clinically suspected allergens was verified in 56 of 78 (71.8%) patients. Agreement with ImmunoCAP varied across platforms: 51.9% for AdvanSure Alloscreen, 61.5% for AdvanSure Alloscreen Max108, 76.7% for SGTi-Allergy Screen (highest), and 50.0% for PROTIA Allergy-Q (lowest) (Table 4). For D. pteronyssinus (n = 10), agreement reached 100% for the SGTi-Allergy Screen, and the concordant positive rate was markedly higher for the SGTi-Allergy Screen (92.9%), highlighting its superior alignment with ImmunoCAP.
Table 4 Concordance with ImmunoCAP Across Four MAST Systems
DiscussionThis single-center retrospective study demonstrated substantial to almost perfect qualitative agreement among four fully automated MAST platforms, with the highest pairwise concordance observed between AdvanSure Alloscreen versus AdvanSure Alloscreen Max108. However, ImmunoCAP verification of the discrepant cases revealed platform-specific differences. Notably, the SGTi-Allergy Screen showed the highest concordance with ImmunoCAP, whereas AdvanSure Alloscreen Max108 provided broader allergen coverage with comparable analytical reliability.
Inter-platform discordance likely reflects differences in antigen sources, extraction and coupling methods, solid-phase chemistry, calibration and class-mapping algorithms, and interference control (eg, cross-reactive carbohydrate determinants, PR-10 proteins). Differences in panel composition (mixed versus single extracts) may further influence class assignments, particularly for plant-derived foods. These assay-level factors are consistent with the larger class discrepancies observed for plant-based allergens (Figure 1) and the platform-specific ImmunoCAP concordance patterns (Table 4).
Our ImmunoCAP concordance was lower than that reported in previous studies.11–13 This can be explained by the design limitations, including partial verification confined to discrepant cases, limited residual serum volume, and small numbers of certain allergens. These factors may have underestimated the concordance. Nevertheless, comparative analyses reflect real-world laboratory practice, where full verification by ImmunoCAP is often impractical owing to cost and workload.
Our findings have practical implications for laboratory adoption and reporting. When alignment with ImmunoCAP is prioritized (eg, confirmatory pathways or component-resolved testing), the SGTi-Allergy Screen may be advantageous Conversely, when broader allergen coverage and operational efficiency are the primary considerations (for example, integrated food-plus-inhalant panels, higher throughput, or minimal sample volume), the AdvanSure Alloscreen Max108 offers a balanced performance. It should be noted that MAST results reflect IgE sensitization and do not by themselves establish clinically relevant allergy; therefore, test results should be interpreted in conjunction with clinical history and symptoms.
This study has some limitations. It was a single-institution, retrospective study that relied on residual serum and applied ImmunoCAP only to discrepant cases, which introduces both spectrum and partial-verification biases. Not all allergens were represented equally, and within-run and between-run reproducibility as well as lot-to-lot variability were not evaluated. Future multicenter prospective studies with larger sample sizes and comprehensive ImmunoCAP verification are warranted to validate these findings and enhance inter-system harmonization, particularly for allergens represented by limited numbers of samples in the present study.
ConclusionsIn summary, the four fully automated MAST platforms showed high qualitative concordance; however, platform-specific differences persisted. Laboratories should balance analytical performance, allergen coverage, and workflow requirements when selecting a system.
AbbreviationsMAST, Multiple Allergen Simultaneous Test; sIgE, Specific Immunoglobulin E; ImmunoCAP (Thermo Fisher Scientific, Uppsala, Sweden; commercial reference assay); κ, Cohen’s kappa coefficient; CPR, Concordant Positivity Rate; PR-10, Pathogenesis-Related protein family 10.
Ethics Approval and Informed ConsentThis study was approved by the Institutional Review Board of Severance Hospital (IRB 2023-3201-002). The requirement for informed consent was waived because the study used residual serum samples obtained during routine clinical testing, posed minimal risk to participants, and involved no additional procedures. All samples and data were anonymized prior to analysis, and no identifiable personal information was accessed. The study was conducted in accordance with the principles of the Declaration of Helsinki.
Consent for PublicationNot applicable. This study does not include any identifiable images, videos, recordings, or personal data.
Author ContributionsAll authors made a significant contribution to the work reported, whether that is in the conception, study design, execution, acquisition of data, analysis and interpretation, or in all these areas; took part in drafting, revising, or critically reviewing the article; have agreed on the journal to which the article has been submitted; reviewed and approved all versions of the article before submission, during revision, and the final version accepted for publication; and agree to be accountable for all aspects of the work.
FundingThere is no funding to report.
DisclosureThe authors declare that Invitros (Cheongju-si, Korea), Sugentech (Cheongju-si, Korea), and Protia (Seoul, Korea) provided the assay reagents in this study. The authors declare no other conflicts of interest in this work.
References1. Brown CR, Higgins KW, Frazer K, et al. Simultaneous determination of total IgE and allergen-specific IgE in serum by the MAST chemiluminescent assay system. Clin Chem. 1985;31:1500–9. doi:10.1093/clinchem/31.9.1500
2. Frati F, Incorvaia C, Cavaliere C, et al. The skin prick test. J Biol Regul Homeost Agents. 2018;32:19–24.
3. Agata H, Yomo A, Hanashiro Y, Muraki T, Kondo N, Orii T. Comparison of the MAST chemiluminescent assay system with RAST and skin tests in allergic children. Ann Allergy. 1993;70:153–157.
4. Rim JH, Park BG, Kim JH, Kim HS. Comparison and clinical utility evaluation of four multiple allergen simultaneous tests including two newly introduced fully automated analyzers. Pract Lab Med. 2016;4:50–61. doi:10.1016/j.plabm.2016.01.002
5. Lee H, Ryu JH, Choi AR, Kim Y, Oh EJ. Inter-laboratory comparison of semi-quantitative allergen-specific immunoglobulin E test: 7 years of experience in Korea. J Clin Lab Anal. 2022;36:e24222. doi:10.1002/jcla.24222
6. Williams P, Sewell WA, Bunn C, Pumphrey R, Read G, Jolles S. Clinical immunology review series: an approach to the use of the immunology laboratory in the diagnosis of clinical allergy. Clin Exp Immunol. 2008;153:10–18. doi:10.1111/j.1365-2249.2008.03695.x
7. Lee YW, Sohn JH, Lee JH, Hong CS, Park JW. Allergen-specific IgE measurement with the IMMULITE 2000 system: intermethod comparison of detection performance for allergen-specific IgE antibodies from Korean allergic patients. Clin Chim Acta. 2009;401:25–32. doi:10.1016/j.cca.2008.10.034
8. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:15974. doi:10.2307/2529310
9. Park SC, Hwang CS, Chung HJ, et al. Geographic and demographic variations of inhalant allergen sensitization in Koreans and non-Koreans. Allergol Int. 2019;68:68–76. doi:10.1016/j.alit.2018.07.005
10. Jeong KY, Park JW. Allergens of regional importance in Korea. Front Allergy. 2021;2:652275. doi:10.3389/falgy.2021.652275
11. Oh EJ, Lee SA, Lim J, Park YJ, Han K, Kim Y. Detection of allergen specific IgE by advansure allergy screen test. Korean J Lab Med. 2010;30:420–431. doi:10.3343/kjlm.2010.30.4.420
12. Lee JH, Park HJ, Park KH, Jeong KY, Park JW. Performance of the PROTIA™ Allergy-Q® system in the detection of Allergen-specific IgE: a comparison with the ImmunoCAP® system. Allergy Asthma Immunol Res. 2015;7:565–572. doi:10.4168/aair.2015.7.6.565
13. Park DJ, Lee J, Kim SY, Kwon HJ, Lee HK, Kim Y. Evaluation of advansure alloscreen max panel with 92 different allergens for detecting Allergen-Specific IgE. Am J Clin Pathol. 2019;151:628–637. doi:10.1093/ajcp/aqz023
Comments (0)